CWP

CWP General Information

Physical Dimensions:
  101 102 103 104 106
Depth (mm) 670 670 670 670 670
Width (mm) 1650 1840 1840 2030 2030
Height (mm) 2010 2010 2010 2010 2010
 
Heating Tank Volume:
  101 102 103 104 106
Liters 260 330 330 330 330
 
Weight:
  101 102 103 104 106
Unit (kg) 460 505 515 555 585
Unit+Packing (kg) 665 755 765 825 855
Operation (kg) 775 905 920 965 1005
 
Pure Water Minimum Outlet Capacity:
  101 102 103 104 106  
+5 °C 8 14 19 23 35  
+10 °C 9 16 21 25 35  
+15 °C 11 18 23 27 35  
+20 °C 13 20 25 29 35  
(in l/min at a pure water outlet pressure of 200 kPA)
 
Water Supply:
  101 102 103 104 106  
Min. input (l/min) 20 30 40 50 60  
Min. input pressure (kPa) 100 100 150 250 350  
Max. input pressure (kPa) 500 (all models)
Temperature Temperature +5 °C to 25 °C
 
Drain Requirements:
  101 102 103 104 106  
Required Capacity (l/min) 20 30 40 50 60  
 
Mains Voltage:
400/230V, 50 Hz, three-phase (five wires) 
415/240V, 50 Hz, three-phase (five wires) 
380/220V, 50 Hz, three-phase (five wires) 
208/120V, 50 Hz, three-phase (five wires) 
Refer to label on the CWP for correct information
 
Power Rating:
RO-Unit 1.8 kW (model 101-102
2.5 kW (model 102-104, 106)
Heating Unit 7.0 kW (model 101)
9.0 kW (model 102-104, 106)
 
Measuring Ranges:
Temperature 0 - 100 °C (±10%)
Flow Inlet Water 3 - 6 l/min (±10%)
Reject Flow 0.2 - 15 l/min (±10%)
Return Flow 0.2 - 20 l/min (±10%) (101-102)
3 - 6 l/min (±10%) (103-104, 106)
Conductivity Inlet Water 100 - 1000 uS/cm (±15%)
Pure Water 2 - 200 uS/cm (±15%)

 

WRO General Information Documents

WATER-10061A CWP Operation with Warm Source Water
In some parts of the country the feed water to the CWP can be a higher than 70 degrees, which is the ideal temperature for the CWP. This update addresses this issue and explains how to apply it to your situation, if you are dealing with temperature issues